Laboratory Furnace upto 1800°C

Laboratory Furnace upto 1800°C with MoSi2 Heating Element
GENERAL FEATURES & DESCRIPTION :
- High Quality MoSi2 1800 or 1900 grade Heating Elements Vertically suspended directly radiating on both sides of the heating cavity to derive symmetrical heat distribution.
- Hot-face Super Refractory Mullite / Bubble Alumina insulation backed by low heat capacity refractory and Fibrous insulations of different grades
- Dual Shell with Fan cooling to ensure low Skin temperature.
- Insert type Swing aside Parallel – moving Door (Chain guided) with Peep Hole facing away from the Operator in any position.
- Fiber wool sealing in the Door for least Heat leakage
- Thyristorised Circuitry with adjustable Current limit and Soft-Start features for maximum protection to Heating Elements and for simple control of heat-up rate.
- Air Cooled Step Down Transformer/s having minimum 3 Nos. Input Step Change facility for desired Secondary Voltage
- Micro-processor based Digital PID Temperature Controller / Programmer with Pt/Rh Thermocouple (Type ‘B’ or ‘LJ’) in pure Alumina Sheath.
- Digital Over-Temperature Protector with ensure overall safety.
- Exhaust air outlet at Roof of the Furnace
- PC Interface for On –line Date Acquisition / Storage Graph / Table also available (Optional)
PERFORMANCE :
- Temperature Control Accuracy within ±1°C of set point.
- Temperature uniformity within ± 5°C near T'max with soaking.
- Carefully Engineered for rugged use and easy maintenance.
- Easy Installation and Hassle Free Operation
APPLICATIONS :
- Trial firing and P.L.C.A.R. Testing
- Industrial R & D work
- Thermal testing in Quality Control Laboratories
- Precision firing of Oxide & Electronic Ceramics and high temperature products.
- Sintering and Processing in Powder Metallurgy
CONSTRUSTION :
- Welded angle-iron frame floor model construction in two sections with Hammer-Tone painted Sheet-metal Paneling.
- Top section comprises the Heating Cavity assembly
- Base frame houses the specially designed Transformer/s complex with Electrical gearing (Analog Voltmeter, Ammeter Switches etc.) thereby needing lesser floor area
- Heat-affectable metallic parts in the vicinity of Door opening are of heat-resisting
- Stainless-Steel fabrication.
CONCISE SPECIFICATIONS :
T' Max |
Model |
Cavity Size (mm) |
Hours to
|
KW |
||
W | H | D | ||||
1600°C
|
60C 9 | 150 | 200 | 300 | 3.0 |
8.0
|
60C 15 | 200 | 250 | 300 | 3.5 | 10.0 | |
60C 42 | 300 | 300 | 450 | 4.0 | 15.0 | |
1700°C
| 70C 5* | 125 | 150 | 250 | 3.0 |
4.0
|
70C 9 | 150 | 200 | 300 | 3.5 |
8.0
|
|
70C 15 | 200 | 250 | 300 | 3.5 | 10.0 | |
70C 42 | 300 | 300 | 450 | 4.5 | 16.0 | |
1800°C |
80C 3* | 100 | 150 | 175 | 4.0 |
3.5
|
80C 4* | 125 | 150 | 215 | 4.0 |
4.5
|
|
80C 15 | 200 | 250 | 300 | 4.5 | 10.0 | |
80C 42 | 300 | 300 | 450 | 5.0 | 16.0 |